Beam bending type strain point and slow cooling point measurement system GM series

Equipment Overview

This is a device for measuring the viscosity of glass in the strain point and slow cooling point regions, and complies with ISO 7884/4 and 7884/7 by employing the beam bending method, which is specified in the JIS standard “JIS R-3103-2, Method for measuring slow cooling point and strain point by fiber stretching method.

Beam bending type strain point and slow cooling point measurement system

Features

  • Automatic operation and data processing by PC.
  • The measuring fixture is made of fused fused silica glass, and errors due to thermal expansion are compensated for by data processing software.
  • Displacement is read under constant load by a servo mechanism.
Specifications
Viscosity Measurement Range Slow Cooling Point, Strain Point, and Range of Viscosity of Glass
Electric Furnace Shape: Twice split front open/close type
Furnace Body Dimensions: 450mm dia. x 350mm H
Inner Dimension of Furnace: φ100×H150mm
Maximum Temperature: 1000°C
Heater: Kanthal AF
Heater Capacity : 2.5KW
Loading mechanism AC servo motor
Load cell Load cell 1Kg
Displacement measurement By pulse position control of servo motor
 Control panel External dimensions: W500×H750×D500mm
 Control method:3-division control byprogram controller
Sequencer: Measurement operation instructions and monitoring by touch panel
PC system Notebook PC, A4 printer, software for measurement operation and data processing

These specifications are subject to change for improvement.

Please feel free to contact us for more information about our glass testers!