Wide Range Viscometer GM Series

Equipment Overview

In the conventional annual measurement in the range of 103 to 1011Poise ( 102 to 1010Pa.s), three different types of equipment had to be used because the measurement principle and measurement systems were spread over three different systems. In addition, it was troublesome to combine the results of each measurement in order to obtain the desired temperature-viscosity characteristic curve.
This device overcomes these problems and streamlines them into a single device.

(*References: Yutaka Shiraishi et al. Journal of the Japan Institute of Metals, Vol. 60, No. 2, 1996 184-191)

Features

  • Viscosity measurement can be continuously performed on the same sample during the temperature increase process.
  • Small sample volume and low temperature error.
  • Automatic measurement is possible, and the data can be printed out as tables and graphs immediately after measurement.

Viscosity measurement range

  • (Whole) 103-1011 poise
  • (Penetration method) 1011 to 109 poise
  • (Parallel plate measurement method) 109-105 poise
  • (Disk rotation method) 105-103 poise
Configuration Dimensions
Measuring instrument and electric furnace stand Dimensions: W550×D570×H1260mm
Electric furnace Shape: Vertical tubular type, split in two, open/close type
Furnace body dimensions: φ450×H225mm
Inner dimensions of furnace: φ80×H100mm
Max. temperature: 1500°C
Heater Capacity: 2.0kW
Measuringsection Dimensions: W570×D470×H1600mm
Torque sensor, Desk rotation device
  Displacement detector: differential transformer ±5mm
  Load detector: Load cell 1Kg
  Torque detector: Torque load cell: 1Kg
  Load servo mechanism: AC servo 50W 0 to 1Kg
  Rotation mechanism: AC servo motor 50W, 0 to 150rpm
Control Unit Dimensions: W570×D470×H1600mm
Sequencer: Touch panel for measurement operation instructions and monitoring
  Temperature program controller, SCR, transformer
Measuring fixture Displacement/torque detection disc, rotary torque detection disc
(platinum disc is optional)
Sample shape 10 mm dia. x 5 mm H
PC system PC, A4 printer, measurement operation and data processing software

These specifications are subject to change for improvement.