Equipment Overview
This system conforms to JIS-R3103-1 “Glass softening point test method” (ISO 7884-6:1987, ASTM C338). The principle of measurement, preparation of samples, etc. are inherited from JIS, and direct reading is automated. The measurement program is operated on a PC, and the softening point measurement results can be confirmed on the PC.

Features
- Measurement program operation is performed on a PC.
- Automatic measurement is performed by using a photo sensor.
- Data can be saved on a PC for direct data processing.
(technical) specification
| Measuring point | Softening point | 
| Furnace dimensions | 150mm dia. x 150mm H | 
| Max. temperature | 800°C | 
| Heater capacity | 450W | 
| Measuring section dimensions | W350×D280×H270mm | 
| Detector | Laser sensor LX2-02 | 
| Accuracy (repeatability) | ±1.5°C | 
| Control panel dimensions | W350×D250×H420mm | 
| Mounted instruments | Displacement amplifier, program controller | 
| PC system | Notebook PC, A4 printer, measurement operation and data processing software | 
These specifications are subject to change for improvement.
Please feel free to contact us for more information about our glass testers!
-185x35.png)